5 maintenance – Eppendorf Research plus User Manual
Page 29

5 Maintenance
29
Eppendorf Research® plus — Operating manual
The values mentioned are theoretical values and are for orientation purposes only. The
volume changes mentioned apply to each volume setting. For the volume change, you
should first set the optimal value for 10% of the nominal volume and then carry out a
gravimetric test. Afterwards carry out gravimetric tests to check 50% and 100% of the
nominal volume with this setting. Change the selected setting again if necessary to achieve
the best possible correction for all volumes. Use the error limits in accordance with ISO
8655-2 (see p. 31) and the technical data provided by Eppendorf AG (see p. 35) to decide
whether the data obtained meets your requirements.
7. If the measured gravimetric values meet your requirements: Slide the locking mechanism up.
Multi-channel
Nominal volume
Color code
+½ revolution
+¼ revolution
–¼ revolution
–½ revolution
10 μL
medium gray
0.53 μl
0.27 μl
–0.27 μL
–0.53 μL
100 μL
yellow
5.4 μl
2.7 μl
–2.7 μL
–5.4 μL
300 μL
orange
10.7 μl
5.4 μl
–5.4 μL
–10.7 μL
6. Slide the locking mechanism up and carry
out a gravimetric test of the changes which
have been made.
5
Main
tenan
ce