5 Maintenance

30

Eppendorf Research® plus — Operating manual

9. Document the changes made and the measurements conducted. Remove the mark on the

volume setting ring and the ejector. The red safety plug on the pipette indicates that the

Research plus has been adjusted and calibrated by the user.

8. Insert the red safety plug on the tool into the

opening of the Research plus and break it

off from the tool.
The red safety plug on the pipette indicates

that the Research plus has been adjusted

by the user. If the adjustment display had

also been set to "0" before, you must close

the opening with a new, red adjustment seal

at the position for the adjustment seal.