5 maintenance – Eppendorf Research plus User Manual
Page 32

5 Maintenance
32
Eppendorf Research® plus — Operating manual
30 - 300 μL
Increment:
0.02 μL
orange
20 - 300 μL
55 mm
30 μL
± 13.0
± 3.9
± 5.0
± 1.5
150 μL
± 3.0
± 3.9
± 1.0
± 1.5
300 μL
± 1.3
± 3.9
± 0.5
± 1.5
100 - 1 000 μL
Increment:
1 μL
blue
50 - 1 000 μL
71 mm
100 μL
± 8.0
± 8.0
± 3.0
± 3.0
500 μL
± 2.0
± 8.0
± 0.6
± 3.0
1 000 μL
± 0.8
± 8.0
± 0.3
± 3.0
0.5 - 5 mL
Increment:
5 μL
purple
0.1 - 5 mL
120 mm
0.5 mL
± 8.0
± 40.0
± 3.0
± 15.0
2.5 mL
± 1.6
± 40.0
± 0.6
± 15.0
5.0 mL
± 0.8
± 40.0
± 0.3
± 15.0
1 - 10 mL
Increment:
10 μL
turquoise
1 – 10 ml
165 mm
1.0 mL
± 6.0
± 60.0
± 3.0
± 30.0
5.0 mL
± 1.2
± 60.0
± 0.6
± 30.0
10.0 mL
± 0.6
± 60.0
± 0.3
± 30.0
Error limits in accordance with ISO 8655-2 - Research plus variable single-channel
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
5
M
a
in
te
nance