5 maintenance – Eppendorf Research plus User Manual
Page 33

5 Maintenance
33
Eppendorf Research® plus — Operating manual
Error limits in accordance with ISO 8655-2 - Research plus variable multi-channel
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
0.5 - 10 μL
Increment:
0.01 μL
medium gray
0.1 - 20 μL
40 mm
0.5 μL
± 48.0
± 0.24
± 32.0
± 0.16
1 μL
± 24.0
± 0.24
± 16.0
± 0.16
5 μL
± 4.8
± 0.24
± 3.2
± 0.16
10 μL
± 2.4
± 0.24
± 1.6
± 0.16
10 - 100 μL
Increment:
0.01 μL
yellow
2 - 200 μL
53 mm
10 μL
± 16.0
± 1.6
± 6.0
± 0.6
50 μL
± 3.2
± 1.6
± 1.2
± 0.6
100 μL
± 1.6
± 1.6
± 0.6
± 0.6
30 - 300 μL
Increment:
0.02 μL
orange
20 - 300 μL
55 mm
30 μL
± 26.0
± 7.8
± 10.0
± 3.0
150 μL
± 5.2
± 7.8
± 2.0
± 3.0
300 μL
± 2.6
± 7.8
± 1.0
± 3.0
5
Main
tenan
ce