beautypg.com

3 research plus fix, 6 technical data, Technical data – Eppendorf Research plus User Manual

Page 38

background image

6 Technical data

38

Eppendorf Research® plus — Operating manual

6.3 Research plus fix

Test conditions according to ISO 8655-1, 8655-2 and 8655-6 for piston-stroke pipettes with air

cushion by means of a fine balance with evaporation protection, tested by the German gauging

office.
Number of determinations: 10; water according to ISO 3696; 20°C to 25°C ±0.5°C constant;

with pre-wetted pipette tip; dispensing against the tube wall.

Research plus fixed volume
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL

10 μL

medium gray
0.1 - 20 μL
40 mm

± 1.2

± 0.12

± 0.6

± 0.06

20 μL

light gray
0.5 - 20 μL L
46 mm

± 0.8

± 0.16

± 0.3

± 0.06

10 μL

yellow
2 - 200 μL
53 mm

± 1.2

± 0.12

± 0.6

± 0.06

20 μL

± 1.0

± 0.2

± 0.3

± 0.06

25 μL

± 1.0

± 0.25

± 0.3

± 0.08

50 μL

± 0.7

± 0.35

± 0.3

± 0.15

100 μL

± 0.6

± 0.6

± 0.2

± 0.2

200 μL

± 0.6

± 1.2

± 0.2

± 0.4

200 μL

blue
50 - 1 000 μL
71 mm

± 0.6

± 1.2

± 0.2

± 0.4

250 μL

± 0.6

± 1.5

± 0.2

± 0.5

500 μL

± 0.6

± 3

± 0.2

± 1

1 000 μL

± 0.6

± 6

± 0.2

± 2

6

Technical data