3 research plus fix, 6 technical data, Technical data – Eppendorf Research plus User Manual
Page 38

6 Technical data
38
Eppendorf Research® plus — Operating manual
6.3 Research plus fix
Test conditions according to ISO 8655-1, 8655-2 and 8655-6 for piston-stroke pipettes with air
cushion by means of a fine balance with evaporation protection, tested by the German gauging
office.
Number of determinations: 10; water according to ISO 3696; 20°C to 25°C ±0.5°C constant;
with pre-wetted pipette tip; dispensing against the tube wall.
Research plus fixed volume
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
10 μL
medium gray
0.1 - 20 μL
40 mm
± 1.2
± 0.12
± 0.6
± 0.06
20 μL
light gray
0.5 - 20 μL L
46 mm
± 0.8
± 0.16
± 0.3
± 0.06
10 μL
yellow
2 - 200 μL
53 mm
± 1.2
± 0.12
± 0.6
± 0.06
20 μL
± 1.0
± 0.2
± 0.3
± 0.06
25 μL
± 1.0
± 0.25
± 0.3
± 0.08
50 μL
± 0.7
± 0.35
± 0.3
± 0.15
100 μL
± 0.6
± 0.6
± 0.2
± 0.2
200 μL
± 0.6
± 1.2
± 0.2
± 0.4
200 μL
blue
50 - 1 000 μL
71 mm
± 0.6
± 1.2
± 0.2
± 0.4
250 μL
± 0.6
± 1.5
± 0.2
± 0.5
500 μL
± 0.6
± 3
± 0.2
± 1
1 000 μL
± 0.6
± 6
± 0.2
± 2
6
Technical data