5 error limits in accordance with iso 8655-2, 5 maintenance, Main tenan ce – Eppendorf Research plus User Manual
Page 31

5 Maintenance
31
Eppendorf Research® plus — Operating manual
5.5.5
Error limits in accordance with ISO 8655-2
Error limits in accordance with ISO 8655-2 - Research plus variable single-channel
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
0.1 - 2.5 μL
Increment:
0.002 μL
dark gray
0.1 - 10 μL
34 mm
0.1 μL
± 100.0
± 0.1
± 75.0
± 0.075
0.25 μL
± 50.0
± 0.125
± 30.0
± 0.075
1.25 μL
± 10.0
± 0.125
± 6.0
± 0.075
2.5 μL
± 5.0
± 0.125
± 3.0
± 0.075
0.5 - 10 μL
Increment:
0.01 μL
medium gray
0.1 - 20 μL
40 mm
0.5 μL
± 24.0
± 0.12
± 16.0
± 0.08
1 μL
± 12.0
± 0.12
± 8.0
± 0.08
5 μL
± 2.4
± 0.12
± 1.6
± 0.08
10 μL
± 1.2
± 0.12
± 0.8
± 0.08
2 - 20 μL
Increment:
0.02 μL
light gray
0.5 - 20 μL L
46 mm
2 μL
± 10.0
± 0.2
± 5.0
± 0.1
10 μL
± 2.0
± 0.2
± 1.0
± 0.1
20 μL
± 1.0
± 0.2
± 0.5
± 0.1
2 - 20 μL
Increment:
0.02 μL
yellow
2 - 200 μL
53 mm
2 μL
± 10.0
± 0.2
± 5.0
± 0.1
10 μL
± 2.0
± 0.2
± 1.0
± 0.1
20 μL
± 1.0
± 0.2
± 0.5
± 0.1
10 - 100 μL
Increment:
0.01 μL
yellow
2 - 200 μL
53 mm
10 μL
± 8.0
± 0.8
± 3.0
± 0.3
50 μL
± 2.0
± 0.8
± 0.6
± 0.3
100 μL
± 0.8
± 0.8
± 0.3
± 0.3
20 - 200 μL
Increment:
0.02 μL
yellow
2 - 200 μL
53 mm
20 μL
± 8.0
± 1.6
± 3.0
± 0.6
100 μL
± 2.0
± 1.6
± 0.6
± 0.6
200 μL
± 0.8
± 1.6
± 0.3
± 0.6
5