Technical data, 1 research plus single-channel variable, 6 technical data – Eppendorf Research plus User Manual
Page 35

6 Technical data
35
Eppendorf Research® plus — Operating manual
6
Technical data
6.1 Research plus single-channel variable
Research plus variable single-channel
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
0.1 - 2.5 μL
Increment:
0.002 μL
dark gray
0.1 - 10 μL
34 mm
0.1 μL
± 48
± 0.048 ± 12
± 0.012
0.25 μL
± 12
± 0.03
± 6
± 0.015
1.25 μL
± 2.5
± 0.031 ± 1.5
± 0.019
2.5 μL
± 1.4
± 0.035 ± 0.7
± 0.018
0.5 - 10 μL
Increment:
0.01 μL
medium gray
0.1 - 20 μL
40 mm
0.5 μL
± 8
± 0.04
± 5
± 0.025
1 μL
± 2.5
± 0.025 ± 1.8
± 0.018
5 μL
± 1.5
± 0.075 ± 0.8
± 0.04
10 μL
± 1.0
± 0.1
± 0.4
± 0.04
2 - 20 μL
Increment:
0.02 μL
light gray
0.5 - 20 μL L
46 mm
2 μL
± 5
± 0.1
± 1.5
± 0.03
10 μL
± 1.2
± 0.12
± 0.6
± 0.06
20 μL
± 1.0
± 0.2
± 0.3
± 0.06
2 - 20 μL
Increment:
0.02 μL
yellow
2 - 200 μL
53 mm
2 μL
± 5
± 0.1
± 1.5
± 0.03
10 μL
± 1.2
± 0.12
± 0.6
± 0.06
20 μL
± 1.0
± 0.2
± 0.3
± 0.06
10 - 100 μL
Increment:
0.1 μL
yellow
2 - 200 μL
53 mm
10 μL
± 3
± 0.3
± 1
± 0.1
50 μL
± 1
± 0.5
± 0.3
± 0.15
100 μL
± 0.8
± 0.8
± 0.2
± 0.2
20 - 200 μL
Increment:
0.2 μL
yellow
2 - 200 μL
53 mm
20 μL
± 2.5
± 0.5
± 0.7
± 0.14
100 μL
± 1
± 1
± 0.3
± 0.3
200 μL
± 0.6
± 1.2
± 0.2
± 0.4
6
Technical data