5 Maintenance

25

Eppendorf Research® plus — Operating manual

5.5.3

Changing the factory adjustment

Auxiliary equipment

Supplied safety plug tool

Supplied pin to loosen the safety plug

The systematic and random errors recorded on delivery can be found in a "Certificate of

Conformity". This certificate is included in the delivery. Changes to the factory adjustment will

render the certificate void.
If the dispensing accuracy is only be to changed temporarily, the correct method is to change the

user adjustment. Before changing the user or factory adjustment, observe the general notes

(see General notes on user and factory adjustment settings on p. 22) and the associated

gravimetric tests.

5.5.4

Research plus variable - changing the factory adjustment

If a gravimetric test indicates that an error needs to be corrected and you are required to carry out

a change to the factory adjustment, proceed as follows:

It is possible to change the factory adjustment with the

corresponding accessories for a Research plus with variable

volume setting. If the factory adjustment of the Research plus has

been changed by a user, this can be recognized by a red safety

plug behind the ejector. If the Research plus has been adjusted and

calibrated by Eppendorf AG, this is indicated by a gray safety plug.

Hint!

In the case of Research plus fixed-volume pipettes, modifications are only possible by

changing the user adjustment (see Changing the user adjustment on p. 24).

1. Check whether the adjustment display on

the side is set to "0".
If the adjustment display is not set to "0",

you will first need to set it to "0" with the

adjustment tool. In this case, instead of

continuing with the factory adjustment

changes, carry out a gravimetric test of the

Research plus with the adjustment display

set to "0".