5 maintenance – Eppendorf Research plus User Manual
Page 28

5 Maintenance
28
Eppendorf Research® plus — Operating manual
This results in the following approximate volume changes:
Single-channel
Nominal volume
Color code
+½ revolution
+¼ revolution
–¼ revolution
–½ revolution
2.5 μL
dark gray
0.106 μl
0.053 μl
–0.053 μL
–0.106 μL
10 μL
medium gray
0.53 μl
0.27 μl
–0.27 μL
–0.53 μL
20 μL
light gray
1.06 μl
0.53 μl
–0.53 μL
–1.06 μL
20 μL
yellow
1.07 μl
0.54 μl
–0.54 μL
–1.07 μL
100 μL
yellow
5.4 μl
2.7 μl
–2.7 μL
–5.4 μL
200 μL
yellow
10.8 μl
5.4 μl
–5.4 μL
–10.8 μL
300 μL
orange
10.7 μl
5.4 μl
–5.4 μL
–10.7 μL
1 000 μL
blue
54 μL
27 μL
–27 μL
–54 μL
5 mL
purple
271 μL
135 μL
–135 μL
–271 μL
10 mL
turquoise
542 μL
271 μL
–271 μL
–542 μL
5
M
a
in
te
nance