Research plus variable multi-channel
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
Random
± %
± μL
± %
± μL
0.5 – 10 μL
Increment:
0.01 μL
medium gray
0.1 - 20 μL
40 mm
0.5 μL
± 12
± 0.06
± 8.0
± 0.04
1 μL
± 8.0
± 0.08
± 5.0
± 0.05
5 μL
± 4.0
± 0.2
± 2.0
± 0.1
10 μL
± 2.0
± 0.2
± 1.0
± 0.1
10 – 100 μL
Increment:
0.1 μL
yellow
2 - 200 μL
53 mm
10 μL
± 3.0
± 0.3
± 2.0
± 0.2
50 μL
± 1.0
± 0.5
± 0.8
± 0.4
100 μL
± 0.8
± 0.8
± 0.3
± 0.3
30 – 300 μL
Increment:
0.2 μL
orange
20 - 300 μL
55 mm
30 μL
± 3.0
± 0.9
± 1.0
± 0.3
150 μL
± 1.0
± 1.5
± 0.5
± 0.75
300 μL
± 0.6
± 1.8
± 0.3
± 0.9
6