Technical data

Multipette

®

/Repeater

®

(X)stream

English (EN)

Test conditions and test analysis in accordance with ISO 8655, part 6. Test with an
analytical balance with a moisture trap which has been inspected by the Office of Weights
and Measures.

• Number of determinations: 10

• Use of water in accordance with ISO 3696

• Inspection at 20 °C – 25 °C ±0.5 °C

• Dispensing onto the tube wall

• Volume tests in the mode Dis

• Speed levels set: 7

Combitip advanced Volume range

Testing
volume

Error limits

Error

Systematic error

Random error

± %

± mL

± %

± mL

10 mL
orange
Increment:
0.01 mL

0.1 mL – 10 mL

1 mL

0.50

0.005

0.25

0.0025

5 mL

0.40

0.02

0.25

0.0125

10 mL

0.40

0.04

0.15

0.015

25 mL
red
Increment:
0.025 mL

0.25 mL – 25 mL

2.5 mL

0.30

0.0075 0.35

0.0088

12.5 mL

0.30

0.0375

0.25

0.0313

25 mL

0.30

0.075

0.15

0.0375

50 mL
light gray
Increment:
0.05 mL

0.5 mL – 50 mL

5 mL

0.30

0.015

0.50

0.025

25 mL

0.30

0.075

0.20

0.05

50 mL

0.30

0.15

0.15

0.075

The test volumes for the systematic and random error of the Multipette
(X)stream/Repeater (X)stream comply with the requirements of ISO 8655, part 5.