Technical data
Multipette
®
/Repeater
®
(X)stream
English (EN)
Test conditions and test analysis in accordance with ISO 8655, part 6. Test with an
analytical balance with a moisture trap which has been inspected by the Office of Weights
and Measures.
• Number of determinations: 10
• Use of water in accordance with ISO 3696
• Inspection at 20 °C – 25 °C ±0.5 °C
• Dispensing onto the tube wall
• Volume tests in the mode Dis
• Speed levels set: 7
Combitip advanced Volume range
Testing
volume
Error limits
Error
Systematic error
Random error
± %
± mL
± %
± mL
10 mL
orange
Increment:
0.01 mL
0.1 mL – 10 mL
1 mL
0.50
0.005
0.25
0.0025
5 mL
0.40
0.02
0.25
0.0125
10 mL
0.40
0.04
0.15
0.015
25 mL
red
Increment:
0.025 mL
0.25 mL – 25 mL
2.5 mL
0.30
0.0075 0.35
0.0088
12.5 mL
0.30
0.0375
0.25
0.0313
25 mL
0.30
0.075
0.15
0.0375
50 mL
light gray
Increment:
0.05 mL
0.5 mL – 50 mL
5 mL
0.30
0.015
0.50
0.025
25 mL
0.30
0.075
0.20
0.05
50 mL
0.30
0.15
0.15
0.075
The test volumes for the systematic and random error of the Multipette
(X)stream/Repeater (X)stream comply with the requirements of ISO 8655, part 5.