37

Troubleshooting

Multipette

®

/Repeater

®

(X)stream

English (EN)

• No battery in the Multipette

stream/Repeater stream.

• Battery is fully discharged

(battery is also discharged
when the device is not used).

Insert battery and charge it for
2 hours.

Battery is defective.

Insert a new battery and charge
it for 2 hours.

Display switches
off during use

• Overcurrent protection has

been triggered.

• Combitip was ejected at the

wrong time.

• High backpressure (viscosity,

blockage) when dispensing.

CAUTION!
The following steps will trigger a
piston movement and may cause
the dispensing of liquid!

1. If the Combitip has been

ejected at the wrong time:
carefully pull cylinder
downward by a few millimeters.

2. Trigger a hardware reset (see

p. 40).

3. Hold Combitip over the tube.

4. Press the reset key.

Display:

Reset key pressed while
dispensing.

Caution!
The following action will trigger a
piston movement and the
dispensing of liquid.

1. Hold the Multipette stream/

Repeater stream over an empty
tube.

2. Press the reset key again.

Display flashes: • Ejector operated during

dispensing, causing the
Combitip cylinder to be
released.

• Cylinder pressed against the

wall too strongly or bent when
dispensing.

• A Combitip cylinder which was

released accidentally is not
inserted or inserted incorrectly.

1. Hold the Multipette stream/

Repeater stream over an empty
tube.

2. To empty the Combitip, press

the reset key again.

3. Press the ejector to remove the

Combitip.

4. Dispose of Combitip.

Symptom/
message

Cause

Remedy

Press reset!

Attention!

Piston moves!

Task aborted!

Press reset to

empty Combitip