8 technical data, Technical data 8.1 – Eppendorf Multipette Xstream User Manual
Page 42

42
Technical data
Multipette
®
/Repeater
®
(X)stream
English (EN)
8
Technical data
8.1
Errors with the Multipette (X)stream/Repeater (X)stream
Combitip advanced Volume range
Testing
volume
Error limits
Error
Systematic error
Random error
± %
± µL
± %
± µL
0.1 mL
white
Increment:
0.1 μL
1 μL – 100 μL
10 μL
1.60
0.16
2.50
0.25
50 μL
1.00
0.50
1.50
0.75
100 μL
1.00
1.00
0.50
0.50
0.2 mL
light blue
Increment:
0.2 μL
2 μL – 200 μL
20 μL
1.30
0.26
1.50
0.30
100 μL
1.00
1.00
1.00
1.00
200 μL
1.00
2.00
0.50
1.00
0.5 mL
purple
Increment:
0.5 μL
5 μL – 500 μL
50 μL
0.90
0.45
0.80
0.40
250 μL
0.90
2.25
0.50
1.25
500 μL
0.90
4.50
0.30
1.50
1 mL
yellow
Increment:
1 μL
10 μL – 1000 μL
100 μL
0.90
0.90
0.55
0.55
500 μL
0.60
3.00
0.30
1.50
1 000 μL
0.60
6.00
0.20
2.00
2.5 mL
green
Increment:
2.5 μL
25 μL – 2 500 μL
250 μL
0.80
2.00
0.45
1.125
1 250 μL
0.50
6.25
0.30
3.75
2 500 μL
0.50
12.5
0.15
3.75
5 mL
blue
Increment:
5 μL
50 μL – 5 000 μL
500 μL
0.80
4.00
0.35
1.75
2 500 μL
0.50
12.50
0.25
6.25
5 000 μL
0.50
25.00
0.15
7.50