Technical data
Multipette

®

/Repeater

®

(X)stream

English (EN)

8

Technical data

8.1

Errors with the Multipette (X)stream/Repeater (X)stream

Combitip advanced Volume range

Testing
volume

Error limits

Error

Systematic error

Random error

± %

± µL

± %

± µL

0.1 mL
white
Increment:
0.1 μL

1 μL – 100 μL

10 μL

1.60

0.16

2.50

0.25

50 μL

1.00

0.50

1.50

0.75

100 μL

1.00

1.00

0.50

0.50

0.2 mL
light blue
Increment:
0.2 μL

2 μL – 200 μL

20 μL

1.30

0.26

1.50

0.30

100 μL

1.00

1.00

1.00

1.00

200 μL

1.00

2.00

0.50

1.00

0.5 mL
purple
Increment:
0.5 μL

5 μL – 500 μL

50 μL

0.90

0.45

0.80

0.40

250 μL

0.90

2.25

0.50

1.25

500 μL

0.90

4.50

0.30

1.50

1 mL
yellow
Increment:
1 μL

10 μL – 1000 μL

100 μL

0.90

0.90

0.55

0.55

500 μL

0.60

3.00

0.30

1.50

1 000 μL

0.60

6.00

0.20

2.00

2.5 mL
green
Increment:
2.5 μL

25 μL – 2 500 μL

250 μL

0.80

2.00

0.45

1.125

1 250 μL

0.50

6.25

0.30

3.75

2 500 μL

0.50

12.5

0.15

3.75

5 mL
blue
Increment:
5 μL

50 μL – 5 000 μL

500 μL

0.80

4.00

0.35

1.75

2 500 μL

0.50

12.50

0.25

6.25

5 000 μL

0.50

25.00

0.15

7.50