beautypg.com

Eppendorf Multipette Xstream User Manual

Page 39

background image

39

Troubleshooting

Multipette

®

/Repeater

®

(X)stream

English (EN)

Further error messages displayed are intended only for error diagnosis by service
personnel. Contact your local Eppendorf partner if the error message cannot be cleared
by pressing the reset key or performing a hardware reset.

Avoiding dispensing errors

The precision and trueness of the Multipette stream/Repeater stream must be verified on
a regular basis. You can find a Standard Operating Procedure (SOP) for the Multipette
stream/Repeater stream on our homepage www.eppendorf.com.

The filled
Combitip should
be emptied or
removed.

• The present residual liquid is

no longer required.

• A different Combitip should be

used.

1. Hold the Multipette stream/

Repeater stream over an empty
tube.

2. To empty the Combitip, press

the reset key.

3. Press the ejector to remove the

Combitip.

4. Dispose of Combitip.

Combitip is
dripping.

• Combitip is not leak tight.

• Combitip is positioned

incorrectly.

Replace Combitip.

Observe the storage conditions.

Battery must be
charged very
often.

• Charging capacity of the

battery has decreased.

• A large number of dispensing

operations has been performed.

If the charging capacity has
decreased:

1. Order a new rechargeable

lithium-ion battery.

2. Insert new rechargeable

lithium-ion battery.

3. Charge new rechargeable

lithium-ion battery for 2 hours
(see p. 20).

Symptom/
message

Cause

Remedy