Danger – IKA HBC 10 control User Manual
Page 35

35
• Cooling coil
When the set temperature close to the ambient temperature, the cooling coil should be used for counter-cooling due
to the self-heating.
Connect the water supply and return line hoses to the cooling coil connectors (see Fig. 2).
For bath temperatures > 95 °C do not use the cooling coil with water.
For bath temperatures > 60 °C make sure that the flow rate through the cooling coil is high enough, to avoid exces-
sive calcification.
• Charging the Battery Pack RB 1 (rechargeable battery)
The battery pack of the WiCo can be charged by any of the following means:
- On the station
- Via a USB cable at the PC or station
- Via an OS 1.0 power supply unit.
1
• Changing battery to WiCo
Please comply with the relevant safety instructions in the "Safety instructions" section for
the RB 1 battery pack!
DANGER
Fig. 13
• WiCo – holder WH 10
WiCo – holder WH 10
WiCo – holder WH 10
Fig. 14
WiCo – holder WH 10
H 44 Boss head clamp
H 38 Holding rod
* Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0,1 g soda (Na
2
CO
3
,