5 database reports, Database reports – BNC SAM 935 Portable Gamma Spectroscopy System User Manual
Page 64
60
SAM 935™ Instruction Manual
7.5
Database Reports
From the U
TIL
menu, it is possible to print a Database Report describing the status of one or all of the
nuclides and their radiation lines. A sample database report is shown here:
This report shows just the Eu152 isotope, but a full database printout would just repeat similar information
for each nuclide in the database. The Enabled column on the left near the nuclide name determines
whether the nuclide as a whole is being processed. The enabled column to the far right shows which radi-
ation lines are enabled for that nuclide. Both must be enabled in order for a line to be processed in the
analysis routines. If the nuclide as a whole is not enabled, individual lines that are enabled will be shown
as (Ena), indicating that they would be enabled if the nuclide was turned on. The efficiency column shows
the efficiencies measured for the standard, or 0.00 if no measurement was ever made. In the latter case,
the efficiency model will be used.
Nuclide Database Printout
Nuclide Halflife Enabled Trig Lvl Energy Intensity Effic SumCorr
Enabled
Eu152 13.542 Y Ena 200U/2k R 41.0 73.80 0.00 4.660 Ena
82.0 0.00 0.00 1.000 Ena
121.8 28.40 0.00 4.600 Ena
162.8 0.00 0.00 1.000 Ena
244.7 7.49 0.00 1.000
344.3 26.60 0.00 1.320 Ena
411.1 2.23 0.00 1.000
444.0 2.78 0.00 1.000
778.9 12.96 0.00 1.210 Ena
867.4 4.15 0.00 1.000
964.1 14.34 0.00 1.000
1101.0 23.68 0.00 1.000 Ena
1408.0 20.85 0.00 1.000 Ena