4 calibration with three coefficients, 3 configure dose rate units and trigger, Calibration with three coefficients – BNC SAM 935 Portable Gamma Spectroscopy System User Manual
Page 21: Configure dose rate units and trigger

SAM 935™ Instruction Manual
17
3.2.4 Calibration with Three Coefficients
To enter the energy calibration factors manually, the factors from a previous live energy calibration for this
unit and matching detector must be available. These numbers can be found on the printout normally done
at the end of a live energy calibration. Alternatively, the numbers can be those recorded earlier from a pre-
vious manual calibration with the three coefficients.
1.
Press the U
TIL
key to bring up the Utilities menu. Use the arrow keys to select
Analysis Tools
Setup. The first screen is the Energy Match Window:
2.
Do NOT change anything in this screen. Press C
ANCEL
(F1) 10 times until you see the question
SET FINE ENGY COEFF?
3.
Press Y
ES
(F2).
The next three screens are standard number entry screens. Change the factors as needed. On
each screen, when the correct value is shown, press E
NTER
.
4.
Return to the normal operating screen.
3.3
Configure Dose Rate Units and Trigger
Because the SAM 935 operates in Surveillance mode by default (also called dose rate monitoring), config-
ure the Dose Rate Units and Dose Rate Trigger options in the Monitor Setup menu:
1.
Press the U
TIL
key to bring up the Utilities menu.