beautypg.com

1 mca presets, Mca presets – BNC SAM 935 Portable Gamma Spectroscopy System User Manual

Page 38

background image

34

SAM 935™ Instruction Manual

5.4.1 MCA Presets

When P

RE

A

DJ

(F4) is selected, the presets are displayed in the data area. When the presets are dis-

played, the F4 function key changes to read HW A

DJ

(or HW I

NFO

when not in Manual mode). The SAM

935 has the following types of preset controls:

ON/OFF

When the preset enable is set to OFF presets will have no effect. It is the
same as setting all presets to zero.

LT Pre

Preset live time.

RT Pre

Preset real time.

Pk Pre

Preset peak count.

Int Pre

Preset integral.

If you are in the manual MCA mode, the presets may be adjusted from this screen. Pressing the

and

arrows moves from preset to preset. The currently selected preset is displayed in the message line. Use
the

and

arrows to change the value of the selected preset.

Note: Preset live and real time will be exact, while in preset peak and integral, the actual
values may exceed the preset slightly.