5 reviewing alarms, Reviewing alarms – BNC SAM 935 Portable Gamma Spectroscopy System User Manual
Page 29
SAM 935™ Instruction Manual
25
4.5
Reviewing Alarms
The SAM 935 can store over 250 alarm events. Each stored alarm consists of a spectrum and all of the
information necessary to analyze the spectrum. Stored alarms can be viewed by using the
←
and
→
arrow
keys:
Review stored alarms by pressing the
←
or
→
arrow keys from the main Surveillance monitoring screen:
• From the main display, press the
←
(left) arrow key to display the previous (most recent) alarm.
• From the main display, press the
→
(right) arrow key to display the first stored (oldest) alarm.
Once you are in review mode, the
←
and
→
arrow keys move from alarm to alarm in a circular fashion as
shown in this example below:
• At any time in the review process, pressing the R
ETURN
(F2) key will bring you back to the main
monitoring screen.
• Press P
RN
S
EL
(F3) to print the selected stored alarm through the serial port. An example of this
printout is shown on page 59, Section 7.1.
• Press MCA (F4) to enter the Manual mode of operation; see Section 5.4 for more information.
A Memory Full message will appear on screen if the stored alarms approach the memory capacity. When
memory is full, the earliest alarms will be discarded to make room for the most recent alarms being stored.
If you see the Memory Full message, you can download the stored alarms to a remote computer, if
desired, and then
Clear Stored Alarms in the Utilities menu.
Main 935 Display
Live (current) alarm
Press
Press
Press
Press
Press
Alarm #2
Press
Alarm #3
Press
Press
Press
Press
(most recent) Alarm #4