SAM 935™ Instruction Manual

43

F2, Edit an Isotope:

Scroll down the list of isotopes to the isotope you wish to monitor. Press
E

NTER

, and you will be returned to the Select list.

F3, Delete a Trigger:

The trigger will be deleted.

F4, Change a Trigger Level: Use the

and

arrows to select a digit; use the

and

arrows to

change the digit. When you’re finished, press E

NTER

.

Monitor and Detail Modes Only: The strength of a trigger is measured in terms of standard deviations
above background, where the background is measured over the same region of the spectrum. For exam-
ple, in the case of a full spectrum, if the number of counts per second in the background spectrum was
100, then its standard deviation would be 10. If the number of counts in a measurement period was 145
counts per second, then the net counts above background would be 145 – 100 = 45, and the reported ratio
would be 45/10 = 4.5. This would trigger an alarm if the corresponding trigger level was set to 4 or less.

6.1.3 Set the Default Trigger Levels

The factory-set default trigger levels are:

5

All Isotopes

8

FULL Spectrum

7

Unidentified Peak

Default values apply only when an item is added to the list of triggers being monitored.

To change the default trigger levels:

1.

Select

Set Default Trigger on the Monitor Setup menu

for the type of default being set (Iso, Full, or Peak) and
press E

NTER

:

2.

Use the

and

arrow keys to select a digit; use the

and

arrow keys to increase or decrease the value.

3.

Press E

NTER

when finished. C

ANCEL

will exit without

changing the original value.

4.

Either select another default trigger level to edit or
press E

NTER

to return to the Monitor Setup menu.

Cursor Points To

F2

F3

F4

I

SOTOPE

– Allows you to

select an isotope

Any Isotope

E

DIT

- Allows one isotope to

be replaced by another

D

ELETE

the FULL trigger

from the list

T

RIG

L

VL

– Allows you to

change the trigger level

FULL

D

ELETE

– Removes the

FULL trigger from the list

T

RIG

L

VL

– Allows you to

change the trigger level

PEAK

D

ELETE

– Removes the

PEAK trigger from the list

T

RIG

L

VL

– Allows you to

change the trigger level