22 [:system:communicate] subsystem, Syst, Comm – BNC 835 Programmer Manual User Manual
Page 63: Unicate, Ubsystem, System:communicate] subsystem

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com
5.22
[:SYSTem:COMMunicate] Subsystem
Command
Parameters
Unit (default)
:SYSTem:COMMunicate:LAN:CONFig
DHCP|MANual|AUTO
DHCP
:SYSTem:COMMunicate:LAN:DEFaults
:SYSTem:COMMunicate:LAN:GATeway
“0.0.0.0”
:SYSTem:COMMunicate:LAN:IP
:SYSTem:COMMunicate:LAN:RESTart
:SYSTem:COMMunicate:LAN:SUBNet
“255.255.255.0”
:LAN:CONFig
:SYSTem:COMMunicate:LAN:CONFig DHCP|MANual|AUTO
:SYSTem:COMMunicate:LAN:CONFig?
This command sets the signal generator’s internet protocol (IP) address.
MANual
The user assigns an IP address to the signal generator.
DHCP
The network assigns an IP address to the signal generator. If DHCP fails, manual
configuration will be used.
AUTO
The network assigns an IP address to the signal generator with a fallback to Auto- IP
if DHCP fails. If both DHCP and Auto- IP fail, manual configuration will be used.
:LAN:DEFaults
:SYSTem:COMMunicate:LAN:DEFaults
This command restores the instrument’s LAN settings to their factory default values.
:LAN:GATeway
:SYSTem:COMMunicate:LAN:GATeway
:SYSTem:COMMunicate:LAN:GATeway?
This command sets the gateway for local area network (LAN) access to the signal generator from
outside the current sub- network. The query returns the current setting, not the saved setting.
:LAN:IP
:SYSTem:COMMunicate:LAN:IP
:SYSTem:COMMunicate:LAN:IP?
This command sets the signal generator’s local area network (LAN) internet protocol (IP) address for
your IP network connection.
:LAN:RESTart
:SYSTem:COMMunicate:LAN:RESTart