BNC 835 Programmer Manual User Manual
Page 38

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com
*RST UP
:LIST:DWELl
[SOURce]:LIST:DWELl
[SOURce]:LIST:DWELl:POINts?
This command sets the dwell time for the current list sweep RAM points. Dwell time is used when
IMMediate is the trigger source. The dwell time is the amount of time the sweep is guaranteed to
pause after setting the frequency and/or power for the current point.
The setting enabled by this command is not affected by signal generator power- on, preset, or *RST.
The variable
:LIST:DELay
[SOURce]:LIST:DELay
[SOURce]:LIST:DELay:POINts?
This command sets the off time for the current list sweep RAM points. Off time is used when
IMMediate is the trigger source. The off time is the amount of time the output is guaranteed to be
blanked after setting the frequency and/or power for the current point to suppress output transients
during a frequency change.
The setting enabled by this command is not affected by signal generator power- on, preset, or *RST.
The variable
:LIST:DELay:AUTO
[SOURce]:LIST:DELay:AUTO ON|OFF|1|0
This command enabled the automatic blanking of the output during transients between two frequency
steps.
*RST ON
:LIST:FREQency:POINts
[SOURce]:LIST:POINts?
This command queries the signal generator for the number of frequency points in the current list
sweep RAM.
:LIST:FREQuency
[SOURce]:LIST:FREQuency
[SOURce]:LIST:FREQuency?
This command sets the frequency values for the current list sweep points.
The maximum number of list sweep points is 3’501.
The variable
The setting enabled by this command is not affected by signal generator power- on, preset, or *RST.
Range Please consult the Data Sheet.