BNC 835 Programmer Manual User Manual
Page 52

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com
[SOURce]:PULM:INTernal:PERiod?
This command sets the pulse period for the internally generated pulse modulation using the variables
If the entered value for the pulse period is equal to or less than the value for the pulse width, the pulse
width changes to a value that is less than the pulse period.
*RST 2.5 ms
Range 200 ns – 10 s
:PULM:INTernal:PWIDth
[SOURce]:PULM:INTernal:PWIDth
[SOURce]:PULM:INTernal:PWIDth?
This command sets the pulse width for the internally generated pulse signal.
This command sets the pulse width for the internally- generated pulse modulation using the variable
If the entered value for the pulse width is equal to or greater than the value for the pulse period, the
pulse width changes to a value that is less than the pulse period.
*RST 1.25 ms
Range 50 nS to PERiod
:PULM:SOURce
[SOURce]:PULM:SOURce INTernal|EXTernal|BITStream
[SOURce]: PULM:SOURce?
If set
*RST INTernal
:PULM:STATe
[SOURce]:PULM:STATe ON|OFF|1|0
[SOURce]:PULM:STATe?
This command enables or disables pulse modulation for the selected path.
*RST 0
:PULM:MODE
[SOURce]:PULM:MODE FIXed | LIST
[SOURce]:PULM:MODE?
This command sets the modulation mode. With “FIXed”, the normal ON/OFF modulation can be
performed. With LIST, an arbitrary bit pattern can be loaded that will be used to modulate the carrier.
LIST mode is available with pulse train option.
*RST FIXed