Eic calibration, Entering operator calibration (mode 1), Setting the implement width – Buhler 535 User Manual
Page 70
435/485/535
3-4
EIC Calibration
Electronic Instrument Control System (EIC) -
Calibration
The EIC system is factory pre-calibrated with the
required operating presets and fault codes, however, it
is necessary to verify Mode calibration settings and
become familiar with the calibration procedure .
To access the Electronic Instrument Control System
for programming and calibration, follow the “Entering
Operator Calibration (Mode )” . All programming and
calibration is done with the key switch in the “RUN”
position and the engine off .
Entering Operator Calibration (Mode 1)
. Hold the SELECT button () for two seconds
2 . The TPM will display “” (2) and “CAL (3) .” (If it
does not, use UP and DOWN arrow button (4) to
change the display to “”)
3 . Hold the SELECT button () for two seconds to enter
into mode (Operator Calibration)
4 . Touch the SELECT button () to cycle through the
six Operator Calibration screens:
1. Setting the implement width
The implement width is used in all area calculations
performed by the TPM . It is a measure of the working
width of the implement .
. Enter the width of the implement . The selected
(flashing) digit (5) can be changed by touching
the UP and DOWN arrow buttons (4) . Other digits may
be selected by touching the LEFT and RIGHT arrow
buttons (6) .
2 . Once the implement width has been entered, touch
the SELECT button () to confirm the change .
Section 3 - Operation
EIC - Calibration
F3-122
SPEED
PTO RPM
AREA
START
STOP
CAL
% SLIP
DISTANCE
SELECT
1
2
3
4
Calibration Screen
Without
TPM
With
TPM
Implement width
Slip percent threshold
-
2
Service alert/intervals
3
3
Area preset
4
4
Fault code display on/off
5
5
English/Metric units selection
6
6
F3-123
SPEED
PTO RPM
AREA
START
STOP
IMP WIDTH CAL
% SLIP
DISTANCE
SELECT
4
5
6
1