Appendix, 1 syringe diameters and rate limits – Next Advance SP1800 User Manual
Page 52
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
47
9/15/04
11. Appendix
11.1 Syringe Diameters and Rate Limits
Manufacturer
Syringe
(cc)
Inside
Diameter
(mm)
Maximum
Rate
(mL/hr)
Minimum
Rate
(µl/hr)
B-D
1
4.699
34.67
0.454
3
8.585
115.7
1.515
5
11.99
225.7
2.954
10
14.43
327
4.279
20
19.05
569.9
7.457
30
21.59
732
9.578
60
26.59
1110
14.53
Monoject
1
5.74
51.74
0.677
3
8.941
125.5
1.643
6
12.7
253.3
3.314
12
15.72
388
5.078
20
20.12
635.7
8.318
35
23.52
868.7
11.37
60
26.64
1114
14.59
Terumo
1
4.7
34.69
0.454
3
8.95
125.7
1.646
5
13
265.4
3.473
10
15.8
392
5.13
20
20.15
637.6
8.343
30
23.1
838
10.97
60
29.7
1385
18.13
Air-Tite
10
15.9
397
5.195
20
20.25
643.9
8.426
30
22.5
795
10.41
50
29
1320
17.28
Syringe
(uL)
Inside
Diameter
(mm)
Maximum
Rate
(ul/hr)
Minimum
Rate
(µl/hr)
Hamilton
Microliter
0.5
0.103
16.66
0.001
1
0.146
33.47
0.001
2
0.206
66.64
0.001
5
0.326
166.9
0.003
SGE
0.5
0.1
15.7
0.001
1
0.15
35.33
0.001