beautypg.com

Next Advance SP1800 User Manual

Page 30

background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1600/1800 Programmable Syringe Pump

Publication #1200-02

25

9/15/04

8.4.2 Example 2: Repeated Dispenses with Suck Back

Dispense 2.0 ml with a 5 minute pause between dispenses. In addition, after each dispense, a volume of
0.25 ml is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep
is sounded to alert the operator to prepare for the next dispense.

Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked
back volume of the previous dispense. By changing the last Phase to a [LP:nn] function, the total number
of dispenses can be set.

When entering a function with associated data, such as with the ‘Pause’ in Phase 5, or the ‘Loop’ in Phase
6, the function is entered in 2 steps. First select the function and store it. Then enter the associated data.

Phase

Function

Rate

Volume

Direction

1

RATE

750 ml/hr

2.0 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

750 ml/hr

0.25 ml

Withdraw

Phase

Function

3

LP:ST

Phase

Function

4

LP:ST

Phase

Function

5

PS:90

Phase

Function

6

LP:03

Phase

Function

7

BEEP

Phase

Function

8

PS:30

Phase

Function

Rate

Volume

Direction

9

RATE

750 ml/hr

2.25 ml

Infuse

Phase

Function

Rate

Volume

Direction

10

RATE

750 ml/hr

0.25 ml

Withdraw

Phase

Function

11

LP:EN

This manual is related to the following products: