Next Advance SP1800 User Manual
Page 30
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
25
9/15/04
8.4.2 Example 2: Repeated Dispenses with Suck Back
Dispense 2.0 ml with a 5 minute pause between dispenses. In addition, after each dispense, a volume of
0.25 ml is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep
is sounded to alert the operator to prepare for the next dispense.
Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked
back volume of the previous dispense. By changing the last Phase to a [LP:nn] function, the total number
of dispenses can be set.
When entering a function with associated data, such as with the ‘Pause’ in Phase 5, or the ‘Loop’ in Phase
6, the function is entered in 2 steps. First select the function and store it. Then enter the associated data.
Phase
Function
Rate
Volume
Direction
1
RATE
750 ml/hr
2.0 ml
Infuse
Phase
Function
Rate
Volume
Direction
2
RATE
750 ml/hr
0.25 ml
Withdraw
Phase
Function
3
LP:ST
Phase
Function
4
LP:ST
Phase
Function
5
PS:90
Phase
Function
6
LP:03
Phase
Function
7
BEEP
Phase
Function
8
PS:30
Phase
Function
Rate
Volume
Direction
9
RATE
750 ml/hr
2.25 ml
Infuse
Phase
Function
Rate
Volume
Direction
10
RATE
750 ml/hr
0.25 ml
Withdraw
Phase
Function
11
LP:EN