5.13 Special Power-Up Functions
The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.
5.13.1 Firmware Version Display
To display the pump’s firmware version, press the left-most arrow key (
/
) while turning on power to
the pump. The display will show:
n.n
, where ‘n.n’ is the firmware version number. Pressing any
key will clear the display.
5.13.2 Reset Pumping Program
To clear out the current Program Function setups, press the right-most arrow key (
) while turning on
power to the pump. The display will show
. Pressing any key will clear the display.
With pumps having as many complex features as the NE-1000 family of pumps, it is easy for a novice
user experimenting with the pump's setup to get the pump into a 'weird' state. Performing this reset
function will bring the pump out of a 'weird' state.
5.13.3 Default Program Pre-Load
Pressing the 'Volume'/'Program Function' key while turning on power to the pump will display the Default
Program Menu and the display will show:
. Currently, the only default program is the
Reciprocating Pump program. When selected, the following program will be pre-set, plus the RS-232
communications mode will be set to Reciprocating Mode. The default program can be modified as
needed. The following default program will be loaded.
The Reciprocating Pump Program sets the pump for use with a second pump, plus cable Part#
CBL-DUAL, to create a continuous infusion system.
Phase
Function
Rate
Volume
Direction
1
RATE
500 ml/hr
10.0 ml
Infuse
Phase
Function
Rate
Volume
Direction
2
RATE
500 ml/hr
10.0 ml
Withdraw
Phase
Function
3
JP:01
5.14 Error and Alarm Messages.
If the value entered is beyond the pumps capabilities or is invalid, or an operational problem occurred, one
of the following error or alarm messages will be displayed:
Pump motor stalled alarm.
Value entered is ‘Out Of Range’.
nn
An out of range error occurred at Pumping Program Phase number ‘nn’, or the value just
entered is out of range.
nn
A Pumping Program error was encountered at Phase number ‘nn’. The indicated Phase is
invalid in the context of the entire Pumping Program.
Key pressed is not currently applicable.