Next Advance SP1800 User Manual
Page 35
![background image](/manuals/466043/35/background.png)
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
30
9/15/04
Phase
Function
6
LP:ST
Phase
Function
7
PS:60
Phase
Function
8
LP:60
Phase
Function
9
LP:05
Phase
Function
10
LP:12
Phase
Function
11
JP:01
8.4.7 Example 7: Sub-Programs
This example shows some of the flexiblities provided by the Program Selection functions. The Pumping
Program starts by refilling the syringe with 50 ml at a fast pumping rate (Phase 1), then the Pumping
Program pauses for user sub-program selection (Phase 3). Then performs 5 dispenses of 10 ml at the
selected rates, then refills the syringe again and pauses for the next user sub-program selection.
The user is given the option of choosing one of three defined sub-programs.
1: Dispense 10 ml at 100 ml/hr (Phase 4)
2: Dispense 10 ml at 500 ml/hr (Phase 7)
3: Dispense 10 ml at 750 ml/hr (Phase 10)
After selecting the sub-program and pressing 'Start', the Pumping program continues execution at the
selected sub-program. After the 10 ml dispense, each sub-program jumps or continues with the loop
counter function (Phase 12). The first 4 loops continues Program Execution with the next user sub-
program selection. After the 5th loop, the program continues with Phase 13, which jumps back to the
syringe refill function and starts the whole program over.
Phase
Function
Rate
Volume
Direction
1
RATE
1500.0 ml/hr
50 ml
Withdraw
Phase
Function
2
LP:ST
Phase
Function
3
PR:IN
Phase
Function
4
PR:01
Phase
Function
Rate
Volume
Direction
5
RATE
100.0 ml/hr
10 ml
Infuse
Phase
Function
6
JP:12
Phase
Function
7
PR:02
Phase
Function
Rate
Volume
Direction
8
RATE
500.0 ml/hr
10 ml
Infuse
Phase
Function
9
JP:12