beautypg.com

8 keypad beep enable, 9 rs-232 pump network configuration – Next Advance SP1800 User Manual

Page 23

background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1600/1800 Programmable Syringe Pump

Publication #1200-02

18

9/15/04

The Reciprocating Pump Program sets the pump for use with a second pump, plus cable Part#
CBL-DUAL, to create a continuous infusion system.

Phase

Function

Rate

Volume

Direction

1

RATE

500 ml/hr

10.0 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

500 ml/hr

10.0 ml

Withdraw

Phase

Function

3

JP:01

."

7.8 Keypad Beep Enable

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, a single short beep will sound whenever a key is pressed on the keypad.

7.9 RS-232 Pump Network Configuration

nn

The pump can be configured to communicate either with a

computer or another pump. Communications with a computer (Address Mode) is the default setting and
will be indicated by the [Ad:nn] display.

When in the default Address Mode, up to 100 pumps can be attached to a computer in a single pump
network. The network address is defined by the 2 digits to the right of the colon (:). The valid range of
addresses is from ‘00’ to ‘99’. If only one pump is attached to the computer, set the network address to 0,
[Ad:00] (factory default).

After the network address is displayed, the baud rate is displayed. Each pump in the pump network and
the computer must have the same baud rate setting. Any arrow key can be used to scroll through the
selection of baud rates. The supported baud rates are: 300, 1200, 2400, 9600, and 19200 (displayed as
[1920]).

To change the communications mode:

With the network address displayed [Ad:nn], press the left-most arrow key to enter the communications
mode menu. Pressing any arrow key will scroll through the menu selections:

Address Mode: Default communications with a computer mode.

Reciprocating Pumps. Sends Start/Stop, Pumping rate and reverse
pumping direction to an attached secondary pump.

Reciprocating Pumps. Sends Start/Stop, Pumping rate and same
pumping direction to an attached secondary pump.

Note:

Communications with a computer requires the accessory cable: CBL-PC-PUMP-7.
Communications with a secondary pump requires the accessory cable: CBL-DUAL

In Reciprocating or Dual Pumps Modes, the secondary pump must be left in its default communications
setting of Address Mode, Address 0, and 19,200 baud rate. See the documentation with CBL-DUAL
cable for more detailed information.

This manual is related to the following products: