8.4.4 Example 4: Complex Dispenses with External Synchronization
A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.
The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.
Phase
Function
Rate
Volume
Direction
1
RATE
750.0 ml/hr
0.5 ml
Infuse
Phase
Function
Rate
Volume
Direction
2
RATE
300.0 ml/hr
1.5 ml
Infuse
Phase
Function
3
BEEP
Phase
Function
4
PS:00
Phase
Function
5
LP:02
Phase
Function
Rate
Volume
Direction
6
RATE
750.0 ml/hr
0.5 ml
Infuse
Phase
Function
Rate
Volume
Direction
7
RATE
300.0 ml/hr
1.5 ml
Infuse
Phase
Function
8
BEEP
Phase
Function
9
LP:ST
Phase
Function
10
PS:60
Phase
Function
Rate
Volume
Direction
11
RATE
500.0 ml/hr
3.75 ml
Infuse
Phase
Function
12
LP:03