Next Advance SP1800 User Manual
Page 20
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
15
9/15/04
Display TTL I/O external connector settings. Press any arrow key to display.
If TTL is selected, the following will be displayed:
aa
Operational Trigger setting. ‘aa’ is current setting.
aa
Directional control setting. 'aa' is current setting.
n
'Pump Motor Operating' TTL output pin configuration. 'n' is
the current setting.
n
*** The "Lockout Disable Key" needs to be inserted to display this setting. ***
Lockout changing settings from the keypad. 'n' is the current setting.
n
Keypad beep enable, where ‘n’ is the current setting.
If standard communications mode with a computer is set, then the following are displayed:
nn
RS-232 pump network address, where ‘nn’ is the network address.
nnnn
RS-232 pump network baud rate, where ‘nnnn’ indicates the baud
rate.
Otherwise, the current communications mode will be displayed:
or
Reciprocating or Dual Pump Communications Mode.
7.1 Power Failure Mode
n
Setting: ‘0’ = Disabled, ‘1’ = Enabled.
When enabled, if the Pumping Program was operating when power to the pump was disrupted, the
Pumping Program will automatically start operating when power is reconnected to the pump. Pressing
any key on the keypad while powering up the pump will stop the Pumping Program from starting.
CAUTION: The Pumping Program will start operating from the beginning of the Pumping
Program (Phase 1), regardless of what part of the Pumping Program was operating when the power was
disrupted.
When the NE-1000 family syringe pump is used as a component in an automated infusion/withdrawal
dispensing system, a Pumping Program can be designed to automatically synchronize the pusher block at
the start of the Pumping Program. This would be accomplished using attached sensors that send signals to
the Pumping Program.