SP1600/1800 Programmable Syringe Pump

Publication #1200-02

41

9/15/04

ALARM SETUP

AL [ ]

Set/query alarm setup mode. Set alarm enables or disables alarm buzzer mode.

POWER FAIL SETUP

PF [ ]

Set/query Power Failure mode. Set Power Failure enables or disables Power Failure mode.

TTL I/O OPERATIONAL TRIGGER SETUP

TRG [ ]

Set/Query TTL I/O Operational Trigger input configuration.

TTL I/O Operational Trigger is set to .

=>

FT

= Foot switch trigger (falling edge)

LE

= TTL Level trigger (rising and falling edge)

ST

= Start only trigger (falling edge)

SP

= Stop only trigger (falling edge)

TTL I/O DIRECTIONAL CONTROL INPUT SETUP

DIN [ 0 | 1 ]

Set/query directional control input setup
Settings => 0 = Falling edge: Infuse, Rising edge Withdraw

Same as "dr:rE" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2
pump continuous infusion system

1 = Falling edge: Withdraw, Rising edge Infuse

Same as "dr:dU" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2
pump dual pumping system

PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION

ROM [ ]

Set/query Pump Motor Operating TTL output configuration (TTL pin 7)

Settings => 0 = Output is logic high only when the pump motor is operating (pumping).

1 = Output is logic high when the pump motor is operating (pumping) or when the

Pumping Program is executing a pause timer.

SET KEYPAD LOCKOUT

LOC [ ]

Set/query keypad lockout mode. Set keypad lockout disables changing any settings from the
keypad unless the "Lockout Disable Key" is inserted.

SET KEY BEEP

BP [ ]

Set/query key beep mode. Set key beep enables or disables key beep mode.

9.4.4 General Control and Status Commands

TTL I/O OUTPUT SETTING

OUT