Next Advance SP1800 User Manual
Page 45
![background image](/manuals/466043/45/background.png)
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
40
9/15/04
Queries volume dispense only. Set not applicable.
Response:
I
Where: “I
withdrawn volume.
CLEAR VOLUME DISPENSED
CLD { INF | WDR }
Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating. Query is not applicable.
INF
= Infusion volume
WDR
= Withdrawn volume
Query is not applicable.
9.4.3 Configuration and Setup Commands
New settings for any of the following commands will be stored in the non-volatile memory.
SET PUMP NETWORK ADDRESS AND BAUD RATE
* ADR [
[ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ][DUAL | RECP]
Set/query pump network address
=>Valid range: 0 to 99
B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The
command response and all further communications will be at the specified baud rate.
DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP
Sets the pump to Reciprocating Pumping mode with a secondary pump.
With both Dual and Reciprocating modes, the baud rate defaults to 19,200.
This is a special system command that will be accepted by the pump regardless of its current address.
Once set, the pump will only respond to commands with the set address and at the specified baud rate.
Note: Once DUAL or RECP is set, the pump will only respond to commands that are preceded by the ‘*’
character. To exit DUAL or RECP mode, reset the address: * ADR 0
Example: *ADR
Query current address setting
*ADR 3
Set pump network address to 3. The pump will now only respond to
commands with address 3
*ADR 5 B 1200
Set the pump network address to 5 and the baud rate to 1200. The
command response will be at 1200 baud
ENABLE SAFE COMMUNICATIONS MODE
SAF [
Set/query Safe communications mode setting.
if