Next Advance SP510 User Manual
Next Advance Pumps
Next Advance
Model: SP1010, SP510 syringe pump
Maximum and minimum flow rates
Manufacturer
Syringe
(cc)
Inside
Diameter
(mm)
Maximum
Rate
(mL/hr)
Minimum
Rate
(µl/hr)
Maximum
Rate
(mL/mn)
B-D
1
4.699
191.1
1.459
3.185
3
8.585
637.9
4.868
10.63
5
11.99
1244
9.495
20.74
10
14.43
1802
13.76
30.04
20
19.05
3141
23.97
52.35
30
21.59
4035
30.79
67.25
60
26.59
6120
46.7
102
Monoject
1
5.74
285.2
2.176
4.753
3
8.941
692
5.28
11.53
6
12.7
1396
10.66
23.26
12
15.72
2139
16.33
35.65
20
20.12
3504
26.74
58.4
35
23.52
4788
36.54
79.81
60
26.64
6143
46.88
102.3
Terumo
1
4.7
191.2
1.459
3.187
3
8.95
693.4
5.291
11.55
5
13
1462
11.17
24.38
10
15.8
2160
16.49
36.01
20
20.15
3514
26.82
58.57
30
23.1
4619
35.25
76.98
60
29.7
7635
58.26
127.2
Air-Tite
10
15.9
2188
16.7
36.47
20
20.25
3549
27.09
59.16
30
22.5
4382
33.44
73.03
50
29
7280
55.55
121.3
Syringe
(uL)
Inside
Diameter
(mm)
Maximum
Rate
(ul/hr)
Minimum
Rate
(µl/hr)
Hamilton
Microliter
0.5
0.103
91.83
0.001
1
0.146
184.5
0.002
2
0.206
367.3
0.003
5
0.326
919.9
0.008
SGE
0.5
0.1
86.56
0.001
1
0.15
194.7
0.002
Hamilton
Micro Liter
10 ul
0.48
1.994
0.016
0.9
SGE
500 ul
3.26
Syringe
(mL)
Inside
Diameter
(mm)
Maximum
Rate
(mL/mn)
Minimum
Rate
(µl/hr)
Air-Tite
140
38
208.3
95.37