Next Advance SP1800 User Manual
Page 43
NEXT>>>ADVANCE
www.nextadvance.com
SP1600/1800 Programmable Syringe Pump
Publication #1200-02
38
9/15/04
Set:
Currently selected Phase is set to
and the requested Phase is recalled from the non-volatile memory. Set is only valid if the
Pumping Program is not operating.
Query response:
Currently selected Phase.
PUMPING PROGRAM FUNCTION
FUN [
Set/query the Pumping Program Phase’s function.
This command is relevant to the currently selected Phase. Set is only valid if the Pumping
Program is not operating
For a more detailed description of Program commands, see sec. 8.3, Program Function
Descriptions.
Rate Data Functions
When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.
RAT
Pumping rate. ‘RATE’
INC
Increment rate. ‘INCR’
DEC
Decrement rate. ‘DECR’
Non-Rate Data Functions
STP
Stop pump. ‘STOP’
JMP
Jump to Program Phase. ‘JP:nn’
LOP
Loop to previous loop start ‘nn’ times. ‘LP:nn’
PRI
Program Selection Input. 'Pr:In'
PRL
Program Selection Label definition. 'Pr:nn'
LPS
Loop starting Phase. ‘LP:ST’
LPE
Loop end Phase. ‘LP:EN’
PAS
Pauses pumping for ‘nn’ seconds. ‘PS:nn’
PAS [n.n]
Pauses pumping for 'n.n' seconds. 'PS:n.n'
IF
If Program input TTL pin low, jump to Phase. ‘IF:nn’
EVN
Set event trigger trap. ‘EV:nn’
EVS
Set event square wave trigger trap. 'ES:nn'
EVR
Event trigger reset. ‘EV:RS’
BEP
Sound short beep. ‘BEEP’
OUT
Set programmable output pin. ‘OUT.n’
PUMPING RATE
RAT [
Set/query pumping rate.
=
l/min
MM
= ml/min
UH
=
l/hr
MH
= ml/hr
Applicable only with “Rate Data Functions”.
When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then
not applicable.