Rs-232 pump network connector wiring – Next Advance SP1070-M User Manual
Page 56
NEXT>>>ADVANCE
www.nextadvance.com
SP1000 series operator’s manual
Manual SP1000 series
49
DIS
Query volume dispensed
CLD { INF | WDR }
Clear volume dispensed
SAF [
Safe communications mode
AL [
Alarm mode
PF [
Power failure mode
TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2 | RL | RH | SL | SH ]
Operational trigger default mode
DIN [ 0 | 1]
Directional input control mode
ROM [0 | 1]
Pump Motor Operating TTL output mode
LOC [ P ] [
Keypad lockout mode or Program Entry Mode lockout
BP [
Key beep mode
OUT 5 { 0 | 1 }
Set TTL output level
IN { 2 | 3 | 4 | 6 }
Query TTL input level
BUZ [ 0 | { 1 [ < n > ] } ]
Buzzer control
VER
Query firmware version
*ADR [
Network address and baud rate (system command, valid
regardless of current address)
*ADR [ DUAL | RECP ]
Set Reciprocating or Dual pumping mode
*RESET
Clears program memory and resets communication
parameters (system command, valid regardless of current
address)
------------------------------------------------------------------------------------------------------------------------------
Network Command Burst =>
- Communicate simultaneously with multiple pumps on a pump network.
Any command preceded by the asterik symbol ‘*’ bypasses the pump network address and Safe mode.
12.2
RS-232 Pump Network Connector Wiring
RS-232
To Network To Computer
PC Com Port Connectors
25 Pin
9 Pin
3 - Receive
2 - Receive
Connect to pump TXD
2 - Transmit
3 - Transmit
Connect to pump RCV
9 - Ground
5 - Ground
Connect to pump GND
RCV from network
TXD to network
TXD to computer
RCV from computer