Next Advance SP1070-M User Manual
Page 33

NEXT>>>ADVANCE
www.nextadvance.com
SP1000 series operator’s manual
Manual SP1000 series
26
Phase
Function
3
LP:ST
Phase
Function
4
LP:ST
Phase
Function
5
PS:90
Phase
Function
6
LP:03
Phase
Function
7
BEEP
Phase
Function
8
PS:30
Phase
Function
Rate
Volume
Direction
9
RATE
750 ml/hr
2.25 ml
Infuse
Phase
Function
Rate
Volume
Direction
10
RATE
750 ml/hr
0.25 ml
Withdraw
Phase
Function
11
LP:EN
9.4.3 Example 3: Ramping the Flow Rate
m
l/h
r
R
at
e
Time
0
150
200
250
Phases
2 - 4
Phases
5 - 8
Phases
9-11
Note: Graph is representative only.
Pumping rate increments and decrements in steps.
Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment
to 250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will
decrement to 150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate
is incremented back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process
is repeated.
Phase
Function
Rate
Volume
Direction
1
RATE
200 ml/hr
0.1 ml
Infuse
Phase
Function
2
LP:ST
Phase
Function
Rate
Volume
Direction
3
INCR
1.0
0.1 ml
Infuse
Phase
Function
4
LP:50
Phase
Function
5
LP:ST