Next Advance SP1070-M User Manual
Page 50
![background image](/manuals/466040/50/background.png)
NEXT>>>ADVANCE
www.nextadvance.com
SP1000 series operator’s manual
Manual SP1000 series
43
SET KEY BEEP
BP [
Set/query key beep mode. Set key beep enables or disables key beep mode.
10.4.4 General Control and Status Commands
TTL I/O OUTPUT SETTING
OUT
Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.
Valid value: 5 (Program Output pin)
Query is not applicable.
TTL INPUT QUERY
IN
Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.
Valid values: 2, 3, 4, and 6.
Query response:
BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]
Sets / queries buzzer
Set:
0 = Turn buzzer off;
1 = Turn buzzer on
if
If
otherwise buzzer beeps
if
Query response: { 0 | 1 }
0 = Buzzer off
1 = Buzzer is on continuously or beeping.
10.4.5 System Commands
SET PUMP NETWORK ADDRESS AND BAUD RATE
* ADR [
[ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ][DUAL | RECP]
Set/query pump network address
=>Valid range: 0 to 99
B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The
command response and all further communications will be at the specified baud rate.
DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP
Sets the pump to Reciprocating Pumping mode with a secondary pump.
With both Dual and Reciprocating modes, the baud rate defaults to 19,200.