Next Advance SP1070-M User Manual
Page 46
![background image](/manuals/466040/46/background.png)
NEXT>>>ADVANCE
www.nextadvance.com
SP1000 series operator’s manual
Manual SP1000 series
39
10.4.1 Program Function Commands
The following commands are relevant to the currently select Program Phase. Note: During a Pumping
Program’s operation, the currently selected Phase can change automatically.
PHASE NUMBER
PHN [
Set/query currently selected Program Phase:
Set:
Currently selected Phase is set to
and the requested Phase is recalled from the non-volatile memory. Set is only valid if the
Pumping Program is not operating.
Query response:
Currently selected Phase.
PUMPING PROGRAM FUNCTION
FUN [
Set/query the Pumping Program Phase’s function.
This command is relevant to the currently selected Phase. Set is only valid if the Pumping
Program is not operating
For a more detailed description of Program commands, see sec. 9.3, Program Function
Descriptions.
Rate Data Functions
When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.
RAT
Pumping rate. ‘RATE’
FIL
Fill syringe to dispensed volume. ‘FILL’
INC
Increment rate. ‘INCR’
DEC
Decrement rate. ‘DECR’
Non-Rate Control Functions
STP
Stop pump. ‘STOP’
PAS
Pauses pumping for ‘nn’ seconds. ‘PS:nn’
PAS
Pauses pumping for 'n.n' seconds. 'PS:n.n'
PRI
Sub-Program Selection Input. 'Pr:In'
PRL
Sub-Program Start Label definition. 'Pr:nn'
LPS
Loop starting Phase. ‘LP:ST’
LPE
Loop end Phase. ‘LP:EN’
LOP
Loop to previous loop start ‘nn’ times. ‘LP:nn’
JMP
Jump to Program Phase. ‘JP:nn’
IF
If Program input TTL pin low, jump to Phase. ‘IF:nn’
EVN
Set event trigger trap. ‘EV:nn’
EVS
Set event square wave trigger trap. 'ES:nn'
EVR
Event trigger reset. ‘EV:RS’
TRG
Override Operational Trigger Default Mode, set to
‘tr:aa’
BEP
Sound short beep. ‘BEEP’
OUT
Set programmable output pin. ‘OUT.n’
PUMPING RATE
RAT [ C | I ] [
Set/query pumping rate.