SP1000 series operator’s manual

Manual SP1000 series

15

7.8

Pump Stalled

When the operation of the motor is impeded due to excessive force needed to drive the syringe, or when
then collar clamp position is reached, the pump will stop, pausing the Pumping Program, and a stall alarm

will occur. The display will show

, the ‘Pumping’ LED will blink, and the buzzer will sound

continuously if alarms are enabled. Also, if the RS-232 Safe Mode is enabled, an auto-alarm message will
be sent to an attached computer.

Pressing any key will stop the buzzer and clear the alarm. When the problem causing the pump motor to
stall has been corrected, the Pumping Program can be resumed from any start trigger source:
‘Start’/’Stop’ key, TTL input, or RS-232 command.

8.

Setup Configuration

To change or view the setup configuration, the Pumping Program must be stopped. Press the
‘Diameter’/‘Setup’ key until the first parameter, ‘PF’ is displayed. After 2 seconds, or when any non-arrow
key is pressed, the next parameter will be displayed (see sec. 6.12, ‘Setup’ Key). Pressing an arrow key
under a value will increment, select, or scroll through the valid values for the parameter. The Setup
Configurations will be displayed in the following order:

n

Power Failure mode, where ‘n’ is the current setting.

n

Low Noise mode, whre ‘n’ is the current setting

n

Alarm mode, where ‘n’ is the current setting.

Display TTL I/O external logic connector settings. Press any arrow key to

select.

If TTL is selected, the following TTL logic settings will be displayed:

aa

Operational Trigger default setting. ‘aa’ is current setting.

aa

Directional control setting. 'aa' is current setting.

n

'Pump Motor Operating' TTL output pin configuration. 'n' is
the current setting.

n

*** The "Lockout Disable Key" needs to be inserted to display this setting. ***
Locks out changing settings from the keypad. 'n' is the current setting.

n

Keypad beep enable, where ‘n’ is the current setting.

If standard communications mode with a computer is set, then the following are displayed:

nn

RS-232 pump network address, where ‘nn’ is the network address.

nnnn

RS-232 pump network baud rate, where ‘nnnn’ indicates the baud rate.

Otherwise, the current communications mode will be displayed:

or

Reciprocating or Dual Pump Communications Mode.