The pump then withdraws 0.25 ml (PH:07), pauses for 1 second (PH:08), then repeats this process if the
Program Input TTL line is low (PH:09), otherwise it continues with the next Phase.
Next, the pump pauses for 10 seconds (PH:10). Then it pauses again for the lesser of another 10 seconds
(PH:12) or until an Event Trigger occurs (PH:11). The Program then restarts (PH:13).
Phase
Function
1
ET:RS
Phase
Function
2
OUT.1
Phase
Function
Rate
Volume
Direction
3
RATE
800.0 ml/hr
5.0 ml
Infuse
Phase
Function
4
OUT.0
Phase
Function
5
ET:07
Phase
Function
Rate
Volume
Direction
6
RATE
800.0 ml/hr
0.0 ml (OFF)
Infuse
Phase
Function
Rate
Volume
Direction
7
RATE
1000.0 ml/hr
0.25 ml
Withdraw
Phase
Function
8
PS:01
Phase
Function
9
IF:07
Phase
Function
10
PS:10
Phase
Function
11
ET:01
Phase
Function
12
PS:10
Phase
Function
13
JP:01
9.4.9 Example 9: Automatic Refill
While a foot switch is held, dispense continuously, starting with a 1.5 mL bolus. When the foot switch is
released, the syringe will refill to the volume that was dispensed.
In the pump’s general setup, set the TTL Trigger to “Foot Switch Hold” mode (TR:FH).
Phase
Function
1
TR:ET
Redirect trigger input to the Event Trap
Phase
Function
2
ET:05
Set Trigger Event Trap to start syringe refill
Phase
Function
Rate
Volume
Direction
3
RATE
1000.0 ml/hr
1.5 mL
Infuse
Begin bolus dispense
Phase
Function
Rate
Volume
Direction
4
RATE
500.0 ml/hr
0.0 ml (OFF)
Infuse
Then infuse continuously
Phase
Function
Rate
5
FILL
1000.0 ml/hr
Refill syringe when foot switch is released