NEXT>>>ADVANCE
www.nextadvance.com
SP1000 series operator’s manual
Manual SP1000 series
31
Phase
Function
4
PR:01
Phase
Function
Rate
Volume
Direction
5
RATE
100.0 ml/hr
10 ml
Infuse
Phase
Function
6
JP:12
Phase
Function
7
PR:02
Phase
Function
Rate
Volume
Direction
8
RATE
500.0 ml/hr
10 ml
Infuse
Phase
Function
9
JP:12
Phase
Function
10
PR:03
Phase
Function
Rate
Volume
Direction
11
RATE
750.0 ml/hr
10 ml
Infuse
Phase
Function
12
LP:05
Phase
Function
13
JP:01
9.4.8 Example 8: Dispensing with Complex Synchronization
:01
Repeat if
Program Input
low
:10
:10
Restart on Event Trigger
or 10 seconds
Event Trigger
W
it
h
d
ra
w
In
fu
se
TTL Program Ouput
TTL High
TTL Low
Event Trigger
Infuse until
Event Trigger
This example demonstrates a complex interaction with external equipment, such as synchronizing with
another syringe pump. The Program includes a variety of interactions with external equipment, which
demonstrates the various control possibilities of the SP1000.
The Pumping Program begins by canceling any previous event traps (PH:01) and raising the ‘Program
Output’ TTL line (PH:02). After 5.0 ml has been dispensed at 800 ml/hr (PH:03), the ‘Program Output’
TTL line is lowered (PH:04), sending a synchronization signal to another device.
The pump then continues to pump at 800 ml/hr (PH:06) until a synchronization signal is received at the
‘Event Trigger’ TTL input, causing the Program to jump to Phase 7 (PH:05).