6.13
Special Power-Up Functions
The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.
6.13.1 Firmware Version Display
To display the pump’s firmware version, press the left-most arrow key (
/
) while turning on power to
the pump. The display will show:
n.n
, where ‘n.n’ is the firmware version number. Pressing any
key will clear the display.
6.13.2 Reset Pumping Program
To clear out the current Program Function setups, press the right-most arrow key (
) while turning on
power to the pump. The display will show
. Pressing any key will clear the display.
With a pump with as many complex features as the SP1000 series pump, it is easy for a novice user
experimenting with the pump's setup to get the pump into a 'weird' state. Performing this reset function
will bring the pump out of a 'weird' state.
6.13.3 Default Program Pre-Load
Pressing the 'Volume'/'Program Function' key while turning on power to the pump will display the Default
Program Menu and the display will show:
. Currently, the only default program is the
Reciprocating Pump program. When selected, the following program will be pre-set, plus the RS-232
communications mode will be set to Reciprocating Mode. The default program will over write any other
program stored in memory and it can then be modified as needed. The following default program will be
loaded and stored in the Pumping Program memory.
The Reciprocating Pump Program sets the pump for use with a second pump, plus cable Part# CBL-DUAL,
to create a continuous infusion system.
Phase
Function
Rate
Volume
Direction
1
RATE
500 ml/hr
10.0 ml
Withdraw
Phase
Function
Rate
Volume
Direction
2
RATE
500 ml/hr
10.0 ml
Infuse
Phase
Function
3
JP:01
6.13.4 Program Entry Mode Lockout
Pressing the ‘Diameter’ key while turning on power to the pump will enter special parameter setup. The
following will be displayed:
n
, currently, the only parameter. The “Program Entry Mode
Lockout”, when enabled, prevents inexperienced users from entering “Program Entry Mode” from the
keypad. Mode Disabled: ‘n’ = 0 (default). Mode Enabled: ‘n’ = 1. When enabled, only the Phase 1
‘Rate’, ‘Volume’ and Pumping Direction can be changed. Cannot be enabled when the Pumping Program
is currently programmed with a multiple Phase Program.
6.14
Error and Alarm Messages
If the value entered is beyond the pump's capabilities or is invalid, or an operational problem occurred, one
of the following error or alarm messages will be displayed:
Pump motor stalled alarm.