SP1000 series operator’s manual

Manual SP1000 series

17

Setting

RS-232 Command Setting

FUN TRG

=>

Name

Function

Ft

0

Foot Switch

Falling edge starts or stops the Pumping Program

FH

1

Foot Switch Hold

Falling edge starts the Pumping Program
Rising edge stops the Pumping Program

F2

2

Foot Switch Reversed

Rising edge starts or stops the Pumping Program

LE

3

Level Control

Falling edge stops the Pumping Program
Rising edge starts the Pumping Program

St

4

Start Only

Falling edge starts the Pumping Program

t2

5

Start Only Reversed

Rising edge starts the Pumping Program

SP

6

Stop Only

Falling edge stops the Pumping Program

P2

7

Stop Only Reversed

Rising edge stops the Pumping Program

rL

8

Start on Low Level

Start Pumping Program on a low level

rH

9

Start on High Level

Start Pumping Program on a high level

SL

10

Stop on Low Level

Stop Pumping Program on a low level

SH

11

Stop on High Level

Stop Pumping Program on a high level

Et

12

Trigger Event

Pumping Program Function Only. Triggers an
event instead of stopping the pump.

8.5

TTL I/O Directional Control Input Configuration

aa

Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping

direction. (See sec. 11.1, TTL I/O Operational Controls). The 2 letter configuration parameter to the right
of the colon (:) is defined as follows:

Setting

Name

Function

rE

Reciprocating Pumps

Falling edge: Infuse; Rising edge: Withdraw

dU

Dual Pump

Falling edge: Withdraw; Rising edge: Infuse

The setting names are relevant to a 2 pump system, whereby the ‘Directional Control Input’ TTL pin is
attached to the second pump’s ‘Pumping Direction Output’ TTL pin.

8.6

Pump Motor Operating TTL Output Configuration

n

Configures the functionality of the 'Pump Motor Operating' TTL output pin (TTL pin 7).

Settings: 0: Sets the output to logic high only when the motor is operating (pumping).

Sets the output to logic low when the motor is not operating or when the Pumping Program is

executing a pause timer or is stopped

1: Sets the output to logic high when the motor is operating (pumping) or when the Pumping

Program is executing a pause timer.
Set the output to logic low when the Pumping Program is stopped

8.7

Keypad Lockout

n

Setting: '0' = Disabled, '1' = Enabled.

*** The "Lockout Disable Key" needs to be inserted into the TTL I/O connector to display this setting***