Universal IS Pressure Transmitter UniTrans

Diagnostics and Service

Subject to change due to technical modifications.

© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany

WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de

146

11

Diagnostics and Service

The following error messages can appear on devices with displays
(see chapter 5.1.3):

12

Disposal

If a failure cannot be repaired, the transmitter must be switched off.The op-
erator then must make sure that it is only switched on again after the failure
has been repaired.
Repairs should only be carried out by the manufacturer. All other repairs or
modifications are unauthorized.

Error Code

Error

Error Correction Measures

E00

ROM-error

Return device to manufacturer

E01

Power supply error

Check power supply

E03

E

2

PROM communications error

Disconnect and reconnect power
supply

E04

Sensor’s temperature range was
exceeded

Return sensor’s temperature to
specified limits

E06

Sensor recognition

Disconnect and reconnect power
supply

E07

General communications error
between the sensor and the con-
trol interface unit

Check the connection between
the sensor and the control inter-
face unit

E08

Error E

2

PROM

send in transmitter for service

Sen Error 0

Sensor recognition

Check the connection between
the sensor and the control inter-
face unit

When storing or disposing the pressure transmitter, take precautions with
regard to remaining media. Remaining media in the pressure port may
be hazardous or toxic!

Dispose of instrument components and packaging materials in
accordance with the respective waste treatment and disposal regulations
of the region or country to which the instrument was supplied.

Attention

Important

This manual is related to the following products: