WIKA IUT-11 User Manual
Page 25
Universal IS Pressure Transmitter UniTrans
Product Description
Subject to change due to technical modifications.
© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany
WIKA Alexander Wiegand GmbH & Co. KG· Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de
22
669
39
.0
5 D
/G
B
06
/2
00
6
100
5.2.3 Functional features of transmitters with HART-Communication
Universal commands
• Displayed unit can be set (mbar, bar, psi, mA, %, m, mm WS ...)
(see chapter 10.2.3)
• Definition of measuring site description and tag number (see 10.2.3)
• Measuring circuit test function (see 10.2.8)
• Temperature and min. / max values can be displayed (see chapter 10.2.3)
• Nominal pressure range of sensor can be displayed (see chapter 10.2.3)
• Cyclic measurement and transmission of the measured value (see chapter 10.2.3)
Common commands
• Calibration of zero point and span (under/without pressure) (see chapter 10.2.4)
• Adjustment of damping/integration of the output signal 0 s ... 40 s
(see chapter 10.2.5)
Device-specific commands
• Inversion of output current signal (see chapter 10.2.5)
• Adjustment of alarm values for the output current (3.6 mA or 21 mA)
(see chapter 10.2.5)
• Adjustment of output signal limits (see chapter 10.2.5)
• Mounting correction of the sensor (see chapter 10.2.7)
• Reset functions (see chapter 10.2.7)
• Password activation (see chapter 10.2.7)
• Setting up a table function for a linearisation of the output signal (see 10.2.6)
• Entering the medium density (see chapter 10.2.6)
• Volume-related units of measured value can be set (l, kg, t, m3, gal, lb) (see 10.2.3)
• Display and graphical illustration of the measuring value over time (see 10.2.10)