Universal IS Pressure Transmitter UniTrans

Transmitter Operation with HART®-functionality

Subject to change due to technical modifications.

© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany

WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de

142

Timer:
• Hrs-TOTAL:

display of overall operating hours

• Hrs-CALIB:

display of operating hours since last calibration

• Hrs-RESET:

display of operating hours since last reset

• Hrs-SENSOR: display of sensor operating hours

Reset:

Via the four buttons certain functions of the transmitter are reset.
• Timer:

Reset of operating hours

• Alarm 4 - 20 mA:

Alarm reset when the 4 ... 20 mA-limit has been exceeded

• MIN/MAX values:

Reset of MIN/MAX-values in the display

• Reset all:

Reset of all settable values to factory settings
(see also chapter 9.4)

Calibrated special measuring ranges i. e. 4 bar together with a 6 bar
transmitter can be adjusted by factory pre-setting. A reset to default will
reset the sensor to its nominal range (i. e. 6 bar). The factory pre-setting
gets lost.