6technical data – WIKA IUT-11 User Manual
Page 28
Universal IS Pressure Transmitter UniTrans
Technical Data
22
669
39
.0
5 D
/G
B
06
/2
00
6
Subject to change due to technical modifications.
© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de
103
6
Technical Data
6.1
Input-values
6.2
Output-values
Pressure Ranges
(Absolute pressure upon request)
/ overpressure limit / Burst pressure
0 ... 0.4 bar
2
2.4
0 ... 1,6 bar
10
12
0 ... 6 bar
35
42
0 ... 16 bar
80
96
0 ... 40 bar
80
400
0 ... 100 bar
200
800
0 ... 250 bar
500
1200
0 ... 600 bar
1200
2400
0 ... 1.000 bar
1500
3000
0 ... 1,600 bar
2000
4000
0 ... 2,500 bar
3000
5000
0 ... 4,000 bar
4400
7000
-1 ... +0 bar*
2
2.4
-1 ... +0,6 bar*
10
12
-1 ... +3 bar*
35
42
-1 ... +5 bar*
35
42
-1 ... +15 bar*
80
96
* only relative pressure
Do not exceed the nominal pressure!
Output signal
4 ... 20 mA, two-wire, optionally with
modulated HART signal
Accuracy [% of span]
(incl. non-linearity, hysteresis, non-
repeatability, zero signal and full
scale error)
< 0.10 at ranges < 1000 bar
< 0.30 at ranges > 1000 bar
Non-linearity
< 0.05% of span (BFSL) per IEC 61298-2
at ranges < 1000 bar
< 0.20% of span at ranges > 1000 bar
Turn down behavior: (1/k)
up to 1 : 5
1 : 5 to 1 : 20
no changes of deviation
the accuracy must be multiplied by the
factor (turn down / 5)
example for TD = 1:15, (k = 15)
accuracy = 0.10 * (15/5) = 0.3