beautypg.com

WIKA HP-2 User Manual

Page 11

background image

11

WIKA operating instructions pressure transmitter, model HP-2

11379180.03 09/2013 GB/D/F/E

GB

6. Commissioning, operation

6. Commissioning, operation

CAUTION!

Prior to commissioning, the pressure transmitter must be subjected to a

visual inspection.

Only use the pressure transmitter if it is in perfect condition with respect to

safety.

6.1 Mechanical mounting

The sealing faces at the instrument have to be

undamaged and clean.

When screwing the instrument in, the force

required to do this must not be applied through

the case, but only through the spanner flats

provided for this purpose and using a suitable

tool.

For the installation point, the valid values for

tightening torque and maximum pressure

should be taken from the high-pressure pipe

supplier's documentation.

When screwing in, do not cross the threads.

For information on tapped holes and welding sockets, see Technical information

IN 00.14 at www.wika.com.

6.2 Electrical mounting

The instrument must be grounded via the process connection.

Use the pressure transmitter with shielded cable, and ground the shield on at least

one end of the lead, if the lines are longer than 30 m (3- or 4-wire) or leave the build-

ing.

Select a cable diameter that matches the cable gland of the plug. Make sure that

the cable gland of the mounted plug has a tight fit and that the seals are present and

undamaged. Tighten the threaded connection and check that the seal is correctly

seated, in order to ensure the ingress protection.

For cable outlets, make sure that no moisture enters at the cable end.

Observe the specifications in accordance with chapter 3 “Specifications”.

Spanner flats

Open-ended spanner

(spanner width 27)