Myron L 729II User Manual
Page 44

40
F. PIEZO ELECTRIC ALARM
-PA
Piezo Electric Alarm ordered with Digital
Monitor/controller.
PAO Piezo Electric Alarm ordered separately for Digital or
remotely mounted Analog Monitor/controller or OEM
models.
1. DESCRIPTION
The PA/PAO is an electronic sound device capable of emitting a
80dB or more @ 30cm, high pitched squeal. See figure III.F.1.
Specifications
Oscillating Frequency — 3.0≠0.5KHz
Operating Voltage (750II) — 24VDC Nom. (1.5-30VDC Max.)
Sound Pressure Level (Min) 30cm/12VDC — 80dB
Current consumption (Max) @ 12VDC — 12mA
Tone — Constant
Operating Temperature — -20 - +60°C
Size — 24 x 9.5 mm
2. INSTALLATION
Briefly -
The PA/PAO Piezo Alarm attaches to the front panel with the tape
supplied, see figure III.F.3.
The wire harness plugs into a methode connector on the main
CB*, see figure III.F.4.
* May also be connected to the SC Module as shown in figure
III.I.5, if desired or so equipped.
WARNING: BEFORE STARTING, IF MONITOR/
CONTROLLER IS INSTALLED, ENSURE THE POWER
IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE
TO THE INSTRUMENT, AND COULD BE HARMFUL OR
FATAL
TO
PERSONNEL.
ONLY
QUALIFIED
PERSONNEL
SHOULD
INSTALL
OR
SERVICE
ELECTRICAL EQUIPMENT.
Physical
NOTE: Remote mounting will require a .25” (6.35mm) hole, and
extending the harness. Use #22 gauge speaker wire. Observe
polarity.
If the front panel has all ready been removed from the enclosure
skip to #4.
1. Using a standard slot screwdriver remove the
screws on the front panel.
2. Carefully wiggle the front panel to loosen the gasket and
pull gently toward you. Do not pull more than about 8
inches/20CM or you could damage the wiring harness.
3. Turn the front panel around so that the back side is
facing you.
4. Using a 1/4” drill or a sharp knife, carefully open the
CENTER front hole in the front panel as shown in figure
III.F.2.
5. Peal off tape backing from PA, and install between
guides as shown in figure III.F.3.
6. Set front panel aside.
7. Connect wire harness to Main CB as shown in figure
III.F.4, or to SC as shown in figure III.F.5.
Test
1. To test, simply turn ON the controller and adjust
controller set point until the alarm sounds off.
NOTE: If the sensor is connected, the solution value, set point
value and Set Point Conversion (SPC) jumpers may affect the
test.
2. Turn power OFF.
!
CAUTION - READ FOLLOWING CAREFULLY
Figure III.F.1
Double Backed Tape
Harness ~9"
Piezo Alarm
Methode Connector
Figure III.F.2
FRONT PANEL
Rear View
CUT OPEN
THIS HOLE